A sensor that won't show up stops an X-ray appointment cold. Before you assume the sensor is dead (and expensive), work through these steps. Most "not detected" problems are a port, a cable or a driver.

Handle the sensor and its cable gently. Don't bend the cable sharply or yank it from the port. Cable damage is a common cause of sensor failures.

Step 1: Try a different USB port, with nothing in between

Anyone can do this

Dentrix's DEXIS troubleshooting guide says to bypass any extension cables and to plug the sensor into several ports, front and back of the PC (Testing and troubleshooting DEXIS/KaVo sensors).

Step 2: Check Device Manager

Anyone can check this

  1. Right-click Start and open Device Manager.
  2. Expand Dental Intraoral X-ray Sensor.
  3. Look for the DEXIS sensor and make sure it says it is working properly.

If the sensor is missing or shows a warning, the drivers are most likely not installed correctly (DEXIS Platinum setup guide). Go to Step 3.

Step 3: Reinstall the driver with DEXIS Software Manager

Needs admin rights on the PC

DEXIS's quick start guide says to download the Intraoral Sensor Driver, extract it, and run the installer as a system administrator. That installs DEXIS Software Manager. Choose the Typical install, then reboot (DEXIS sensor quick start).

DEXIS notes that Software Manager doesn't apply to DTX Studio users or some third-party imaging software (DEXIS sensor jump start). Call your IT provider or DEXIS first if you aren't sure which imaging software your office runs, or if you don't have administrator rights on the PC.

Problems like this often point to a bigger gap. Get your free HIPAA technical score → 62 questions, about 10–15 minutes, with your top 3 gaps and a PDF.

Step 4: Check the calibration files (older sensors)

IT or DEXIS support job

On DEXIS Platinum sensors using legacy drivers, a calibration file's name must match the serial number printed on the sensor's cable, near the USB plug. If the sensor was swapped or the PC was replaced, the right file may be missing. DEXIS can supply calibration files by serial number (setup guide).

Newer Ti2 and IXS sensors calibrate on their own when you start an acquisition or connect the sensor in the imaging software (quick start), so this step usually doesn't apply to them.

Step 5: Read the DEXIS log

Anyone can check this

DEXIS writes errors to C:\ProgramData\DEXIS Imaging\Polaris\Log. The Dentrix troubleshooting guide explains the common codes (source):

Step 6: Test it in your practice software

Anyone can do this

If you use Dentrix, the same guide walks through a test exposure: choose Bitewing Single Image, pick the teeth, set the acquisition method to DEXIS/KaVo Sensors, then Proceed with Acquisition. It notes the sensor must already have worked in the DEXIS imaging software on that PC.

Prevent sensor problems before they happen

Which imaging software are you using?

The fix can depend on the software that captures the image. DEXIS notes that DTX Studio users and some third-party imaging software users don't use DEXIS Software Manager for drivers (jump start guide). If your office captures through Dentrix, Eaglesoft or another bridge, the bridge settings also need to point at the right sensor. If you don't know which one you use, look at the program that opens when you take an X-ray, and tell support its name.

Still not working?

DEXIS technical support is 888-883-3947, Monday to Friday, 9am to 6pm Eastern (DEXIS contact page). Have the sensor's serial number and the error codes from the log ready.

In Metro Detroit, call us at (313) 516-4114. We set up and support imaging PCs, sensors and the servers behind them for dental offices. See our dental IT support. If the whole imaging system feels sluggish, read why dental software gets slow. Using a Schick sensor instead? Read Schick sensor not detected.

Common questions

Why is my DEXIS sensor not detected?

Most "not detected" problems are a USB port, an extension cable or a driver. Plug the sensor straight into the PC, try other ports including the back, then check Device Manager. A missing or warning entry usually means the driver needs reinstalling.

Can I plug a DEXIS sensor into a USB hub or extension cable?

Bypass extensions while you troubleshoot. If your office needs one, use the powered, active USB extension or hub that DEXIS sells, not a generic one.

What do DEXIS error codes 7, 9, 10 and 70 mean?

Error 70 means no sensor attached and error 7 means detector current too high: bypass extensions and try other ports first. Errors 9 and 10 (dark image too high or too low) point to the sensor itself, so contact DEXIS.

What is the DEXIS support phone number?

DEXIS technical support is 888-883-3947, Monday to Friday, 9am to 6pm Eastern (dexis.com, checked November 2026). Have the sensor serial number and any log error codes ready. In Metro Detroit you can also call SpeedyTech at (313) 516-4114.

Still not detecting the DEXIS sensor?

Call us, or send the serial number and any DEXIS log error. We'll tell you what to try next, or when it's time to call DEXIS. A real person at SpeedyTech in Clinton Township reads every message and gets back to you by phone or email. You can also use the chat in the corner.

Have this ready:

Metro Detroit (Macomb and Oakland County). Dental imaging and operatory PCs → Before you reinstall imaging software on a PC that holds images, call first.

About the author. Mike Siemen owns SpeedyTech MSP and Speedy Tech Computer Repair on Hayes Rd in Clinton Township. His team supports dental, medical and small-business offices across Macomb and Oakland County. More about Mike and the team