The patient is in the chair, the sensor is in your hand, and the software says it can't find it. Before you assume a Schick sensor is dead, work through the checks below. Most of the checks are quick, and every step comes from Dentsply Sirona, Patterson or Henry Schein One documentation, linked as you go.
Which driver does your office use? Most of these steps assume the current Schick driver, called IOSS (Intraoral Sensor Software). Dentsply Sirona's current download is IOSS 3.2, and support for the old Schick Driver 5.16, the AE Interface driver and CDR Intraoral TWAIN ended on December 31, 2024 (Dentsply Sirona: Schick software). Eaglesoft 24.20 and newer uses IOSS. Eaglesoft 24.00 and older uses the legacy driver (Patterson KB 20595). For Dentrix, Dentsply Sirona recommends moving to IOSS 3.2, as Dentrix's help notes (Dentrix: Schick/Sirona sensors setup overview). One exception on the same page: an older Schick CDR 2000 sensor may still need the old driver, because IOSS isn't compatible with it. Not sure which one you have? Your IT provider can tell you in a minute.
Step 1: Check the lights on the USB interface
Anyone can check this
The USB 3.0 interface is the small box between the sensor cable and the computer, used with Schick 33 and Schick AE sensors. It has two lights. Here is what the interface status light means, per the Schick 33 USB interface and sensors user guide:
- Off: the cable is not connected. Go to Step 2.
- Amber: no connection to the imaging software is detected. Go to Step 4.
- Green: connected to the imaging software. Slowly "breathing" green means ready to take an X-ray. If it's ready and the software still can't see the sensor, go to Step 5.
If the second light, the sensor status light, is amber, the guide says there is an error condition, such as an incompatible sensor. The USB 3.0 interface works only with Schick AE and Schick 33 sensors.
Older blue-cable USB 2.0 interface? Its lights mean different things, so check the USB 2.0 user guide instead of the list above.
Step 2: Restart, reconnect, and try a rear USB port
Anyone can do this
Henry Schein One's "device not found" checklist starts with these (Dentrix: Installing Intraoral Sensor Software):
- Restart the computer and try again.
- Disconnect the sensor, reconnect it, and try again.
- Try a different USB port, preferably one on the back of the computer that is connected directly to the motherboard.
Hubs and extensions: Dentsply Sirona says the AE USB interface can connect to the PC directly or through an active USB hub that has its own power supply, not one powered by the USB port (Schick AE operating instructions). If a cheap unpowered hub or extension is in the chain, take it out while you test. With the older USB 2.0 interface, the user guide says chained Schick cables should not be longer than 15 feet in total, and that a replaceable cable that isn't attached properly can cause an intermittent connection and keep the sensor from working properly, so check that its screws are tight (Schick USB and sensors user guide).
Handle the cable gently. Don't kink it, pinch it in a drawer, or pull the plug out by the cable.
Step 3: Look for the sensor in Device Manager
Anyone can check this
Right-click Start and open Device Manager. Patterson's troubleshooting article tells you where the sensor should be (Patterson KB 44814):
- USB 3.0 interface: under Universal Serial Bus controllers, listed as Schick AE. (Dentrix's help calls the entry "AE Sensor".)
- USB 2.0 interface: under CDR DEVICES, listed as CDR Elite.
If it isn't there, or it shows a yellow warning, Windows isn't seeing the interface. That points to the cable, the port or the driver. Try Step 2 again with a different port. If that doesn't work, the driver is the next suspect (Step 7).
Step 4: Make sure the Sirona sensor service is running
Needs admin rights on the PC
IOSS runs a Windows service called Sirona Intraoral Sensor Service. Patterson's setup article says it should be set to Automatic (Delayed Start), logged on as Local System, and set to restart on every failure (Patterson KB 44313).
- Press Windows+R, type
services.mscand press Enter. - Find Sirona Intraoral Sensor Service. If it isn't running, right-click it and choose Start.
Stuck on an hourglass? The service is set to Automatic (Delayed Start) on purpose. It starts about 2 minutes after Windows' other automatic services, which Patterson says lets everything it needs load first so the sensor works reliably after a restart (KB 44313). If Eaglesoft shows an hourglass when the sensor is plugged in, Patterson's fix is to close Advanced Imaging, restart the Sirona Intraoral Sensor Service, and confirm it is set to Automatic (Delayed) (KB 44814).
Step 5: Refresh the sensor in the Configuration App
Anyone can do this
IOSS comes with the Intraoral Sensor Configuration Application. Open it from Start > Intraoral Sensor Software (Patterson KB 44814). The Dentrix checklist continues (source):
- Click the Sensors tab, then Refresh Sensors.
- If the sensor shows up, select it and read its status.
- Run a sensor or USB interface test. You may need to close the acquisition agent first so the sensor isn't locked.
Another program can lock the sensor. Henry Schein One notes that a "device not found" error can appear when another imaging program has the sensor locked. Close other imaging software on that PC and try again.
Step 6: Check the settings in Eaglesoft or Dentrix
Anyone can check this
Eaglesoft: go to File > Preferences > X-ray. Default Device should be Sensor and Default Sensor should be Schick sensor (Patterson KB 44313). With IOSS on Eaglesoft 24.20 and newer, auto-detect is built into IOSS, not set in Eaglesoft. Note that IOSS doesn't support taking X-rays through a Terminal Services or Remote Desktop session (same article).
Dentrix Imaging Center: when you start the exam, the Acquisition method for this imaging exam list should be set to Sirona/Schick USB 3/AE, then click Proceed with Acquisition (Dentrix help). The first time a sensor is used on a PC, Dentrix may say correction files could not be found while they install from the sensor. The help page says this is expected.
Step 7: Drivers, firmware and Windows updates
IT provider or vendor support job
If the steps above didn't find it, the cause is usually on the driver side. These fixes involve uninstalling software and updating firmware, so call your IT provider or vendor support before you start:
- Two drivers fighting. Henry Schein One says to confirm IOSS is installed and uninstall any other Schick drivers on the computer. Going the other way, Patterson says IOSS must be uninstalled before the legacy drivers will work (KB 20595). One exception: on PCs that hold images taken before Eaglesoft 24.20, Patterson's IOSS setup also installs parts of the CDR 5.16 driver so the older images display correctly (KB 44313).
- Firmware. Patterson says to update the USB module firmware with only the module plugged in, then update the sensor firmware (KB 44814). For the USB 3.0 module update, Patterson recommends stopping the Sirona Intraoral Sensor Service first (KB 44274).
- It broke right after Windows updates. Windows updates have broken Schick sensors before. Patterson published a fix for Eaglesoft 24.20 and newer offices on IOSS after the October 14, 2025 Windows updates (KB 44883). Don't remove updates on your own. Call your IT provider.
- A checkerboard image instead of an X-ray: Patterson lists a newer-design cable or a repair return as the fix (KB 44814). Call Patterson or Dentsply Sirona.
Step 8: Still not working? Who to call
- Eaglesoft offices: Patterson support, 800-475-5036, 7am to 7pm Central, Monday to Friday (Patterson KB 5786).
- Dentrix offices: Dentrix support, 800-336-8749, Monday to Thursday 6am to 6pm and Friday 6am to 5pm Mountain Time (Dentrix support page). If the sensor test fails in the Configuration App, Dentrix's help says to contact Sirona support.
- The sensor itself: the Schick 33 user guide lists Dentsply Sirona Technical Service for U.S. customers at 800-659-5977.
Have the sensor's serial number, the interface light color, what Device Manager shows, and any error message ready.
In Metro Detroit, call us at (313) 516-4114. We set up and support the imaging PCs, sensors and servers behind them for dental offices. See our dental IT support. Using DEXIS instead? Read DEXIS sensor not recognized.
Prevent it next time
- Keep every X-ray PC on the same driver and version. If one room runs IOSS and another runs the legacy driver, a sensor that works in one room may not work in the next.
- Plan for the USB 2.0 interface. Dentsply Sirona says the USB 2.0 interface is discontinued but stays in support through July 2027 (source). If you still use one, budget for the change now instead of finding out mid-appointment.
- Install the software before you plug it in. On a new or replaced PC, the Schick 33 guide says not to connect the USB interface until the setup program has run successfully.
- Test after big Windows updates. Take a quick test image at the start of the day after a major update, before the first patient.
- Write down serial numbers. Keep a list of each sensor's serial number, which room it lives in, and its warranty date. Support will ask for it.
Common questions
Why is my Schick sensor not detected?
On the USB 3.0 interface, check the status light: off means the cable isn't connected, amber means no connection to the imaging software, and breathing green means ready. (The older USB 2.0 interface uses different lights.) Then restart the PC, reconnect, try a rear USB port, and check that the Sirona Intraoral Sensor Service is running. Another imaging program holding the sensor can also cause a device not found error.
What is IOSS?
IOSS (Intraoral Sensor Software) is Dentsply Sirona's current Schick sensor driver. Eaglesoft 24.20 and newer uses it, and Dentsply Sirona recommends IOSS 3.2, as Dentrix's help notes. Dentsply Sirona ended support for the older Schick Driver 5.16 on December 31, 2024. An old CDR 2000 sensor may still need the old driver.
Can I plug a Schick sensor into a USB hub?
Dentsply Sirona says the Schick AE USB interface can connect directly or through an active USB hub with its own power supply, not a bus-powered hub. While troubleshooting, plug it straight into a port on the back of the PC.
Why does Eaglesoft show an hourglass when the Schick sensor is plugged in?
Patterson's fix is to close Advanced Imaging, restart the Sirona Intraoral Sensor Service, and confirm it is set to Automatic (Delayed Start). The delayed start lets the other Windows services load first so the sensor works reliably after a restart.
Still not seeing the Schick sensor?
Call us, or send the color of the interface lights, what Device Manager shows, and whether you use Eaglesoft or Dentrix. A real person at SpeedyTech in Clinton Township reads every message and gets back to you by phone or email. You can also use the chat in the corner.
Have this ready:
- What you are seeing (a phone photo of any error is fine)
- How many computers or people it affects
- Anything that changed recently
Metro Detroit (Macomb and Oakland County). Dental imaging and operatory PCs → Before you update firmware or reinstall imaging software, call first.